Thursday, April 4, 2013

Aliasing on Doppler Ultrasound

The main forms of Doppler used pulse wave Doppler(PWD), Color Doppler (CD) and Power Doppler (PD). Modern ultrasound equipment with CD felicities usually offers conventionally B-mode real-time scanning. CD, which shown color representation of blood flow towards and away from probe, pulsed wave spectral analysis shown as waveforms above and below a baseline and often PD.

The Doppler waveforms is constructed from a series of samples. For a given sampling rate or pulse repetition frequency (PRF) , there is a maximum frequency. Above this , the sampled waveform cannot be constructed accurately and a lower frequency signal is produced.

The sample Doppler signal adequately , at least two samples must be taken on the shortest cycle for the correct frequency to be determined. This is called Nyquist limit. This can al,so be expressed as the sample rate being at least 2 x maximum frequency in the Doppler signal. If the sample rate, controlled by a PRF, is less than the essential twice per cycle, then a interpretation of the high frequencies above one half 0f the sampling frequency. Because the signals is not sampled quickly enough, it appears below the baseline. The PRF therefore should be at least twice the maximum values of Doppler frequency to be measured in order to avoid aliasing.

Saturday, March 23, 2013

Color Doppler Flow and Displacement Imaging


Ultrasonic B-mode real-time imaging can be combined with Doppler in a scanner so that the scanner is capable of providing not only anatomical information but also blood flow data. Both sets of information are displayed simultaneously. A cursor line is typically superimposed on the B-mode image to indicate the direction of the Doppler beam. An FFT algorithm is used to compute the Doppler spectrum that is displayed in real time. This type of scanner is called a duplex scanner; a duplex image is shown in Figure below. Alternatively, blood flow data can be displayed in real time and superimposed with the B-mode image if the data acquisition rate and image processing algorithms are sufficiently fast.

ultrasoud
where the magnitude and phase of H(τ) are an even function and an odd function, respectively. The symbol A is used to represent the magnitude of H(τ) here. where the dot operation represents the first derivative=∂H(τ)/∂τ. Let< ω > denote the mean of ω and, from the definition of mean angular frequency and Equation However, limited by the frame rate and field of view, the pulse repetition frequency in most color Doppler systems is between 8 and 16 kHz. As a result, aliasing frequently occurs with color Doppler in cardiac imaging. To overcome these
problems, one can reduce the image size or use M-mode color Doppler, in which the beam is fixed in one direction. In the heart, the myocardium is in motion during a cardiac cycle. Tissue color
Doppler images of this motion can be acquired with color Doppler methods previously described as well. The difference lies in that myocardial motion is slower than blood flow and myocardial echoes are stronger than blood. The spurious Doppler signals from blood in this case can be eliminated by thresh holding the echoes. A tissue Doppler image of the heart in which the color indicates the velocity of myocardial motion is shown in Figure 6.5. Many clinical applications have been found for color Doppler flow imaging, including diagnosing tiny shunts in the heart wall and valvular regurgitation and stenosis. It considerably reduces the examination time in many diseases associated with flow disturbance. Problematic regions can be quickly identified first from the flow mapping. More quantitative conventional Doppler measurements are then made on these areas. Color Doppler has been now widely used in a variety of medical disciplines; however, it has several shortcomings:

  • Flow perpendicular to the beam cannot be reliably detected.
  • Higher blood flow velocity results in aliasing.
  • Its spatial resolution is poorer than B-mode gray-scale imaging.
  • The mean velocity estimated is the average velocity within a pixel or voxel.
  • Because the color Doppler image is overlaid over the gray-scale B-mode, the overlay process is determined arbitrarily by thresh holding, which may result in vessel wall overwrite obscuring the slow blood flow signal near the wall.
  • Large echoes due to slow moving tissues can cause the “color flash artifact” because they overlap echoes from flowing blood.

Tuesday, March 19, 2013

Color Duplex Imaging Limitation

As a result of the vast amount of information to be processed, color duplex scanning has a much poorer spatial and temporal resolution than pure B-mode imaging. Axial resolution is proportional to the wavelength in B-mode imaging while it is dependent on the number of sample volumes placed along the color Doppler scan line in the color duplex mode. The use of smaller sample volumes improves axial resolution but at the expense of sensitivity and accuracy in Doppler shift evaluation as the signal-to-noise ratio deteriorates. Lateral resolution in color-coded duplex scanning is determined by the number of color Doppler lines processed per centimeter. The frame rate decreases as the number of Doppler lines increases, resulting in a lower temporal resolution, in particular at greater scan depths. As a result of these limitations, the axial resolution of color duplex ultrasound on the order of 0.4–1.0 mm with a lateral resolution of only 1.0–2.0 mm, which is 4–10 times lower than B-scan resolution (Widder 1995). The frame rate in the color duplex mode ranges from 50–200 ms (corresponding to an image repetition rate of 5 Hz), depending on the scanning depth and size of the color box. When a low pulse repetition frequency is selected, the speed at which the color Doppler scan lines sweep the sector is similar to or slightly below the mean flow velocity in arteries. Therefore, a single ultrasound scan may simultaneously depict systolic flow (e.g. displayed in red) and early diastolic flow(e.g. displayed in blue) . Due to the low temporal resolution, however, the color coding does not fully reflect the pulsatile character of flow. Slow flow produces smaller Doppler frequency shifts, which have to be extracted from short echo pulse packets for each scan line consisting of a number of individual pulses. The scan lines must be processed successively. Though the insonation angle should ideally be as small as possible for optimal Doppler scanning, this is not always practical because there will be a longer delay when the color box is tilted (beam steering). This is why one must find a compromise, in particular when examining deeper vessels. Tilting the color box 20° and 30° prolongs the echo arrival time by 13% and 31%, respectively.

Color duplex imaging, like all diagnostic ultrasound techniques, is impaired by scattering and acoustic shadowing caused by bowel gas or calcified structures (bone or calcified plaques on vessel walls). The examiner can circumvent such interfering structures by moving the transducer, but this is frequently achieved only at the cost of a longer echo arrival time due to a greater distance from the structure of interest. Strong reflectors that are oblique to the beam axis act like mirrors and generate phantom images in another area of the scan. Such mirror images can be identified by angling the transducer, which will make the mirror artifacts disappear or appear in a different location. When the ultrasound beam strikes interfaces of high acoustic impedance at a right angle, reverberations (repeat echoes) may occur with the ultrasound pulses being reflected to and fro, resulting in a kind of ping-pong effect. Slight angulation of the transducer prevents reverberations but will also reduce reflection from the interface and thus degrade image quality.

Color Doppler, Tissue Doppler & Multigate Doppler

Tissue Doppler & Multigate Doppler

In the Doppler signal processing chain  , if only large amplitude echoes of lower Doppler frequencies from tissues such as myocardium or heart muscle are retained and low amplitude echoes of higher Doppler frequencies are suppressed, the motion of the tissues can be monitored. This will be touched upon again in the next chapter. An amplitude threshold can be set to allow only the larger echoes to pass through. Tissue Doppler has been proven a clinically useful tool for assessing the state of myocardium. In conventional pulsed Doppler, only one gate is used to measure blood flow within the sampled window or sampling volume confined by the beam width and the gate duration. If blood flow velocities at multiple points along the ultrasound beam need

Doppler Flow Measurements to be measured, pulsed Doppler flow meters with multiple gates (e.g., 8 or 16 gates) have been developed. These devices allow the measurement of velocities in real time across the lumen and thus have been used frequently to determine the blood flow velocity profile in arteries.

Color Doppler

A new way of displaying color Doppler information, i.e., “power mode” or “energy mode” imaging, has been introduced to minimize some of the color Doppler problems (Rubin et al., 1994; Zagzebski, 1996) and has been well accepted by the clinical community. All top-of-the-line scanners now have this option. Instead of the mean Doppler shift, the power contained in the Doppler signal is displayed in this approach. There are several advantages to doing so:

• A threshold can be set to minimize the effect of noise.

• The data can be averaged to achieve a better signal-to-noise ratio.

• The images are less dependent upon the Doppler angle.

• Aliasing is no longer a problem because only the power is detected. As a result, signals from blood flowing in much smaller vessels can be detected. The images so produced have an appearance similar to that of x-ray angiography.